Home Telephone Location
logo
Call us on: +44 (0) 28 9077 3030
Banner

Multiparameter Meters

Hanna HI-2600 Series EC/TDS/Salinity with logging

HI-2631 EC, TDS & Salinity Benchtop Meter with Data Logging

The HI-2631 Digital Benchtop Meter is designed for fast, accurate measurements of electrical conductivity (EC), total dissolved solids (TDS), salinity, and temperature. It utilises the Hanna Instruments HI-763100 digital conductivity electrode with integrated temperature sensor, providing reliable, temperature-compensated results for laboratory, classroom, and industrial applications.

In addition to EC and TDS, the HI-2631 supports salinity measurement and calibration and features built-in data logging, enabling easy record keeping, USB transfer, and traceable analysis. Automatic probe recognition simplifies setup and reduces errors, while the intuitive interface makes operation effortless.

Ideal for laboratories, research facilities, classrooms, and industrial quality control, the HI-2631 combines precision measurement, ease of use, and long-term reliability.

Key Features

  • EC / Conductivity measurement

  • TDS (Total Dissolved Solids)

  • Salinity measurement and calibration

  • Temperature measurement

  • Data logging and USB export

  • Automatic probe recognition

EC - using HI-763100 four-ring conductivity probe    
   EC Range   0.00 to 29.99 µS/cm;
30.0 to 299.9 µS/cm;
300 to 2999 µS/cm;
3.00 to 29.99 mS/cm;
30.0 to 200.0 mS/cm;
up to 500.0 mS/cm,
and absolute conductivity (TDS) is the conductivity value without temperature compensation
 Resolution   0.01 µS/cm;
0.1 µS/cm;
1 µS/cm;
0.01 mS/cm;
0.1 mS/cm
 Accuracy (@25°C)   ±1% of reading (±0.5 µS/cm or 1 digit, whichever is greater)
 Calibration   Cell factor calibration; six standards available: 84µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 118.8 mS/cm, one point offset: 0.00 µS/cm
   Temperature Compensation   Automatic, -5 to 100.0°C; NoTC, can be selected to measure absolute conductivity
 Temperature Coefficient   0.00 to 6.00%/°C (for EC and TDS only), default value is 1.90%/°C
TDS - using HI-763100 four-ring conductivity probe    
   TDS Range   0.00 to 14.99 mg/L (ppm);
15.0 to 149.9 mg/L (ppm);
150 to 1499 mg/L (ppm);
1.50 to 14.99 g/L;
15.0 to 100.0 g/L;
up to 400.0 g/L absolute TDS using 0.80 conversion factor
 Resolution   0.01 mg/L (ppm);
0.1 mg/L (ppm);
1 (ppm);
0.01 g/L;
0.1 g/L
 Accuracy (@25°C)   ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
 Calibration   through EC calibration
 TDS Factor in Standard Mode   0.40 to 0.80 (default value is 0.50)
Salinity - using HI-763100 four-ring conductivity probe    
   Salinity Range   0.0 to 400.0 % NaCl;
2.00 to 42.00 PSU;
0.00 to 80.00 g/L
 Resolution   0.1 % NaCl;
0.01 PSU;
0.01 g/L
 Accuracy (@25°C)   ±1% of reading
 Calibration % NaCl only   one point with HI-7037 standard
Temperature    
   Temperature Range   -20.0 to 120.0 °C
 Resolution   0.1°C
 Accuracy   ±0.5°C
Additional Specifications    
Logging   Up to 1000 (standard mode) (400) records organised in:
• Manual log-on demand, maximum 200 logs
• Manual log-on-stability, maximum 200 logs
• Interval logging in Standard Mode, maximum 600 logs (100 lots)
PC Interface   USB-C
Storage Interface   USB-C
Power supply   USB Type C (5 VDC; 500mA)
Environment   0 to 50°C; Maximum 95% RH non-condensing
Dimensions   205 x 160 x 77 mm
Weight   Approximately 0.85 kg

Share with a friend

Sign up to our monthly Promotions Letter